Specifications
| F-900 Specifications | |
|---|---|
| Display | Sunlight visible transflective graphic LCD |
| Operating environment | 0°C - 45°C (0-90% humidity non-condensing) |
| Battery Capacity | 5 hours - Rechargeable Li-Ion (5000 mAh) |
| Dimensions | 183.5mm x 111 mm x 120 mm |
| Weight | 2.96 kg |
| Enclosure | Anodized aluminum |
| Air Sampling Rate | 80 mL/min |
| Measuring Rate | 1 second intervals, open or closed loop |
| Warm-up time | 1 minute |
| C₂H₄ PPB Sensor | Electrochemical |
|---|---|
| Nominal Range | 0- 10 ppm |
| Resolution | 0.001 ppm |
| Lower Detection Limit | 0.025 ppm (25 ppb) |
| Accuracy | 5% ± 25 ppb |
| Offset Recalibration | Weekly |
| Span Recalibration | *3 months (calibration gas required) |
| C₂H₄ PPM Sensor | Electrochemical |
|---|---|
| Nominal Range | 0- 200 ppm |
| Resolution | 0.1 ppm |
| Lower Detection Limit | 0.5 ppm |
| Accuracy | ±5 % ±0.5 ppm |
| Offset Recalibration | weekly |
| Span Recalibration | *3 months (calibration gas required) |
typically 5% drift/month
| GC Emulation Mode Specifications | |
|---|---|
| C₂H₄ Range | 0.8-20 ppm |
| Accuracy | ±10 % |
| Minimum Sample Volume | 5 mL |
Optional Sensors
| CO₂ PPM Sensor | Low-Power Non-Dispersive Infrared Gas Analyzer |
|---|---|
| Chopping Frequency | 1Hz |
| Source Life | 5000 Hours |
| Nominal Range | 1 – 2000 ppm (Standard) |
| Resolution | 0.1 ppm |
| Repeatability | ±0.1 ppm (short term) |
| Accuracy | 5% |
| Sample Cell | 100 mm x 10.2 mm (3.94" x 0.4" diameter) |
| Warm-up time | <3 minutes |
| Weight | 172.6 g |
| Offset Recalibration | Weekly |
| Span Recalibration | 6 months (calibration gas required) |
| CO₂ Percent Sensor | Infrared Sensor, Pyroelectric detector |
|---|---|
| Nominal Range | 0 – 20% |
| Resolution | 0.01% |
| Accuracy | 5% |
| Offset recalibration | 6 months |
| Span recalibration | 6 months (calibration gas required) |
| Lifetime | >5 years |
| Warm-up time | <3 minutes |
| Weight | 7 g |
| O₂ Percent Sensor | Electrochemical |
|---|---|
| Nominal Range | 0-100% |
| Resolution | 0.1% |
| Offset recalibration | 6 months |
| Span recalibration | 6 months (calibration gas required) |
| Operating environment | 0°C - 55°C (5-95% humidity non-condensing) |
| Weight | 7 g |