Setup Factory Settings

The Setup > Factory Settings menu is used to back up or restore the factory defaults for the calibration parameters for all of the sensors (C2H4, CO2 and O2) on the device. The calibration parameters can also be modified using this menu. It is recommended to always back up the current parameters before making any modifications.

To modify calibration parameters:

  1. Back up the current parameters (Setup>factory settings>back up)
  2. Make the modification in the .cfg fil
  3. Import the .cfg file (Setup>factory settings>restore) to apply the changes