Setup RH Conversion
The Setup RH Conversion menu is used to correct for the temperature sensor being inside F-950 housing. This causes the temperature sensor to reflect the temperature of the F-950 and not the ambient temperature. The temperature is usually several degrees above ambient. The Relative Humidity is calculated based on temperature.
Enter the correct ambient temperature in degrees Celsius, measured from an external temperature sensor, to have the F-950 use for calculating the relative humidity (RH %). To use the entered temperature for RH, “use sample T” should be set to “Yes.”
The current (internal RH) and corrected (sample RH) relative humidity are displayed at the bottom of this screen.
RH Conversion
Enter Sample T(C) 20.0
Use Sample T Yes
(for measurements)
Internal RH (%) 43.6
Sample RH (%) 21.3